You are here:

Introduction | Workflow Requirements

Workflow Requirements

Supports Illumina TruSight Tumor 170 assay
Compatible with NextSeq or HiSeq 2500 Rapid Run systems
Sequencing settings:
2 x 101 (recommended)
Paired End (required)
Connectivity to send sequencing run output data to BaseSpace Sequence Hub; it is recommended to stream run data directly to BaseSpace from the sequencing instrument
Additional specifications can be found in the TruSight Tumor 170 data sheet
Sample IDs on sample sheets configured using the Sample Sheet Template or the BaseSpace® Sequence Hub Prep Tab must meet the following criteria:
< 60 characters
Alphanumeric, underscore, and dash characters only
No duplicate Sample ID
Underscores and dashes must be preceded and followed by an alphanumeric character (no strings of multiple underscores or dashes, sample names cannot begin or end with an underscore or dash)
Sequencing Run Setup
Optional for NextSeq, Import CSV template for BaseSpace® Sequence Hub Prep Tab
Required for HiSeq 2500 Rapid Run, Sample Sheet Template
For DNA analysis
Minimum of 80 M reads (40 M sequencing clusters) is recommended
Paired-end samples with the same read lengths is required
For RNA analysis
Minimum of 16 M reads is recommended
Paired-end samples with the same read lengths are required

For Research Use Only. Not for use in diagnostic procedures. 

©2017 Illumina, Inc. All rights reserved.