putativefusions.csv
The putativefusions.csv file is generated from fusion calls based on the Balance5p3p and ExpressionControl target types.
Two Balance5p3p targets are designed for the driver gene, one on the 3′ end and one on the 5′ end. Reads aligned to these targets and the ExpressionControl targets are used to calculate the imbalance described below. The Imbalance score reflects the magnitude of 5′/3′ imbalance for the targeted driver gene. A higher imbalance score may suggest a potential fusion event involving that driver gene, however additional testing would be necessary to confirm the partner gene and fusion break point. Currently, there are Balance5p3p targets available for four driver genes (ALK, RET, ROS1, NTRK1), all contained in the Focus Panel.
The file contains the following columns.
|
Column |
Description |
|
Gene |
Gene symbol of the targeted fusion partner |
|
Transcript ID |
Transcript ID of the targeted fusion partner |
|
Target ID 5′ |
Target ID of the amplicon on the 5′ of the targeted fusion partner |
|
Target ID 3′ |
Target ID of the amplicon on the 3′ of the targeted fusion partner |
|
Exon 5′ |
Exon(s) of the amplicon on the 5′ of the targeted fusion partner |
|
Exon 3′ |
Exon(s) of the amplicon on the 3′ of the targeted fusion partner |
|
Supporting Reads 5′ |
Read count of the 5′ target of the gene. |
|
Supporting Reads 3′ |
Read count of the 3′ target of the gene. |
|
Expression Control Reads |
Number of reads aligned to targets of type ExpressionControl. |
|
Imbalance |
Calculated by the following equation:
|