Bcl Files

The *.bcl files can be found in the BaseCalls directory:

<run directory>/Data/Intensities/BaseCalls/L<lane>/C<cycle>.1

They are named as follows:

s_<lane>_<tile>.bcl

The *.bcl files are binary base call files with the format described below.

Bytes

Description

Data type

Bytes 0–3

Number N of cluster

Unsigned 32bits little endian integer

Bytes 4–(N+3)

Where N is the cluster index

Bits 0-1 are the bases, respectively [A, C, G, T] for [0, 1, 2, 3]:
bits 2-7 are shifted by two bits and contain the quality score.

All bits ‘0’ in a byte is reserved for no-call.

Unsigned 8bits integer

 

 