Callability Report

DRAGEN appends the percent callability for each region included in the BED file specified by the test and generates a full-callability-report in both BED and CSV formats. Callability is defined as the fraction of non-N reference positions having a PASSing genotype call. The following criteria are used to calculate callability metrics:

• Callability is calculated over the gVCF.
• Multi-allelic variants are not counted.
• Decoy contigs are ignored.
• Unplaced and unlocalized contigs are ignored.
• N positions are considered non-callable.
• For regions where no variant calling was performed, callability is 0.
• A homozygous deletion counts as a PASSing genotype call for all the reference positions spanned by the deletion.