MiSeq Reporter FAQs

Collapse All

|

Expand All

  • Workflow


  • When MiSeq Reporter analysis is finished, a checkmark appears in the State column of the Analyses tab and the CompletedJobInfo.xml file is written to the root level of the analysis folder.

    MiSeq Reporter software runs as a web server that runs continuously on the MiSeq computer. To view current MiSeq Reporter status, open your web browser and enter the following URL: http://PCname:8042 (where “PC Name” is the name or IP address of the MiSeq PC).

    If a new sequencing run is started on the MiSeq System before secondary analysis of a previous run is complete, secondary analysis will be stopped automatically. MiSeq computing resources are dedicated to either sequencing or analysis, and the system is designed in such a way that a sequencing command overrides an analysis command. Secondary analysis can later be requeued from the MiSeq Reporter Analyses tab.

    MiSeq Reporter needs to have access to the repository, which is the location of the folder containing data for analysis. You can set this location in the settings window on the MiSeq Reporter main screen. When MiSeq Reporter has access to the repository, your runs appear in the Analyses tab in MiSeq Reporter. Select the Requeue checkbox next to the run you want to analyze, and then click Requeue to start analysis.

    For a 2 × 250 bp run, analysis takes about 3 hours when using the latest PC RAM configuration on the MiSeq. Genome size for resequencing also affects analysis time. If analysis is taking longer than two hours, consider mapping to a more appropriate reference for your sample, or perform analysis offline by installing MiSeq Reporter on another computer.

    If an alignment is performed against the whole genome, then the analysis time will be significantly longer than two hours. Also, bioinformatics analysis for metagenomics may take as long as 12 hours.

    No, secondary analysis cannot be delayed automatically. Secondary analysis will automatically stop if you begin the run setup steps for a subsequent sequencing run on the MiSeq System.

  • File Formats


  • FASTQ files are generated during secondary analysis by MiSeq Reporter for most analysis workflows. To generate only FASTQ files, specify the GenerateFASTQ workflow in the sample sheet, which generates FASTQ files and then exits secondary analysis.