How to thaw and store sequencing reagents for optimal performance

03/11/22


Sequencing reagent performance depends on thawing technique and thawed reagent storage time and conditions. It is important to understand and follow the recommendations for reagent thawing and storage because some reagents contain enzymes that are sensitive to temperature and time. Unless specified below, refreezing thawed reagents can negatively affect enzyme activity and run performance.

iSeq 100 | MiniSeq | MiSeq | NextSeq 500/550 | NextSeq 1000/2000 | HiSeq 1000/1500/2000/2500 | HiSeq 3000/4000/X | NovaSeq

iSeq 100 Reagent Kits

    After receiving reagent shipment, store reagent cartridge in the freezer at -20 to -25°C with the arrow pointing up for at least 24 hours before thawing for use. There are several ways to thaw the reagent cartridge:
    • If thawed in a room temperature water bath (20 to 25°C) for a minimum of 6 hours, the cartridge is stable up to 18 hours in the water bath
    • If thawed in a refrigerator (2°C to 8°C) for a minimum of 36 hours, the cartridge is stable up to 72 hours in the refrigerator
    • If thawed in room temperature air (20°C to 25°C) for a minimum of 9 hours, the cartridge is stable up to 18 hours in room temperature air
    • iSeq 100 i1 Reagent v2 kits can be thawed at 4°C for 36 hours and are stable at 4°C for up to one week prior to use

MiniSeq Reagent Kits

    There are several ways to thaw the reagent cartridge:
    • If thawed in a 37°C water bath for 35 minutes, it is stable up to 2 hours
    • If thawed in a room temperature water bath (19°C to 25°C) for 90 minutes, it is stable up to 24 hours
    • If thawed at room temperature air (19°C to 25°C) for 5 hours, it is stable up to 24 hours
    • If thawed at 2°C to 8°C for 18 hours, it is stable up to 72 hours

MiSeq v2 and v3 Reagent Kits

    There are several ways to thaw the reagent cartridge:
    • If thawed in a room temperature deionized water bath for 60 minutes, place on ice or set aside at 2°C to 8°C and use within 6 hours
    • If thawed overnight at 2° to 8°C, store at 2° to 8°C for up to 1 week

NextSeq 500/550 Reagent Kits

    There are several ways to thaw the reagent cartridge:
    • If thawed in a room temperature deionized water bath for 60 minutes, place on ice or set aside at 2°C to 8°C, and use as quickly as possible or within the same day
    • If thawed overnight at 2° to 8°C, store at 2° to 8°C for up to 1 week

NextSeq 1000/2000 Reagent Kits

    There are several ways to thaw the reagent cartridge, remove the cartridge from box and proceed to thaw the cartridge in foil seal bag with bag label facing up:
    • Water bath: Thaw in temperature controlled 25°C water bath, with a minimum of 9.5 cm water depth for 6-8 hours.
    • Refrigerator (2° to 8°C): One day prior to anticipated run, remove cartridge from -25°C to -15°C storage. Thaw at room temperature for 6 hours. Transfer to refrigerator 2° to 8°C, and continue to thaw for a minimum of 12 hours. The cartridge is stable up to 72 hours in the refrigerator. When thawing at room temperature, cartridge should have air on all sides except the bottom and should not be stacked. In the refrigerator, cartridges may be stacked but should be given ample space on all other sides.
    • Room temperature (20°C to 25°C) for a minimum of 9-16 hours.
    Transfer cartridge to 4°C after thawing, cartridge is stable for 72 hours at 4°C. Cartridges stored at 4°C will need to be brought to room temperature for 15 minutes to 1 hour prior to run start. There is no easy way to determine if the cartridge is fully thawed; so it is important to follow these guidelines as well as plan ahead. Thawing times are long.

HiSeq 1000/1500/2000/2500 Cluster and SBS Kits

TruSeq SR Cluster Kit v3, TruSeq PE Cluster Kit v3, HiSeq SR Cluster Kit v4, and HiSeq PE Cluster Kit v4

    There are several ways to thaw the reagent plate:
    • If the cBot reagent plate is thawed in a room temperature water bath for 60 minutes, use within the same day
    • If the cBot reagent plate is thawed at 2°C to 8°C overnight, no longer than 16 hours, use the plate within the same day
    Store the thawed clustering kit on ice or at 2°C to 8°C until use. For best results, use as soon as possible after thawing.

TruSeq Rapid v1 Cluster Kit and HiSeq Rapid v2 Cluster Kit

    There are several ways to thaw the reagent plate:
    • If the cBot reagent plate is thawed in a room temperature water bath for 30 minutes, use within the same day
    • If the cBot reagent is plate thawed at 2°C to 8°C overnight, no longer than 16 hours, use the plate within the same day
    Store the thawed clustering kit on ice or at 2°C to 8°C until use. For best results, use as soon as possible after thawing.

HiSeq v4 High Output SBS Reagents and HiSeq Rapid v2 SBS Reagents

    There are several ways to thaw the reagents:
    • If the reagents are thawed in a room temperature deionized water bath for about 90 minutes, use within the same day
    • If the reagents are thawed at 2°C to 8°C overnight, use the next day
    Store thawed CRM on ice in a separate container to avoid cross-contamination. Store the thawed SBS reagents on ice or at 2°C to 8°C until use.

HiSeq 3000/4000/X Cluster and SBS Kits

HiSeq 3000/4000 SR Cluster Kit, HiSeq 3000/4000 PE Cluster Kit, and HiSeq X Five/Ten Cluster Kit

  • Thaw the cBot reagent plate in a room temperature water bath for 60 minutes
  • Thaw the EPX1, EPX2, EPX3, and RSP reagents at room temperature for 10 minutes, then set on ice
    For best results, use as soon as possible within the same day. Prepare the ExAmp reaction master mix immediately before use and load onto the cBot within 15 minutes of preparation.

HiSeq 3000/4000 SBS Reagents and HiSeq X Five/Ten SBS Reagents

    There are several ways to thaw the reagents:
    • If the reagents are thawed in a room temperature deionized water bath for about 90 minutes, use within the same day
    • If the reagents are thawed at 2°C to 8°C overnight, use the next day
    Store thawed HCM on ice in a separate container to avoid cross-contamination. Store thawed SBS reagents on ice or at 2°C to 8°C until use.

NovaSeq Reagent Kits

NovaSeq Cluster and SBS cartridges

  • Thaw SP, S1, and S2 cluster cartridges in a room temperature deionized water bath for 2 hours
  • Thaw S4 cluster cartridges in a room temperature water bath for 4 hours
  • Thaw SP, S1, S2, and S4 SBS cartridges in a room temperature water bath for 4 hours
    Store thawed reagents at 2°C to 8°C until use for up to 24 hours. Cluster and SBS cartridges can be refrozen once at -20°C if run cannot be set up right away.

NovaSeq XP Workflow Reagents

  • Thaw DPX1, DPX2 and DPX3 at room temperature for 10 minutes, then set aside on ice
    For best results, use as soon as possible. For the best performance, prepare the ExAmp master mix immediately before use. Store the ExAmp master mix up to 1 hour on ice, or 30 minutes at room temperature. After loading the libraries and ExAmp, load flow cells onto the instrument within 30 minutes. If ExAmp reagents cannot be used, they can be refrozen one time only. If you must refreeze, do so immediately after thawing.