SNV Calling
The variant caller filters the set of filtered and realigned reads for SNV calling without affecting indel calls. Any contiguous trailing sequence of N base calls is trimmed from the end of the read. Using a mismatch density filter, reads with an unexpectedly high number of disagreements with the reference are masked as follows.
|
•
|
The variant caller identifies each insertion or deletion as a single mismatch. |
|
•
|
Base calls with more than 2 mismatches to the reference sequence within 20 bases of the call are ignored. |
|
•
|
If the call occurs within the first or last 20 bases of a read, the mismatch limit is applied to a 41-base window at the corresponding end of the read. |
|
•
|
The mismatch limit is applied to the entire read when the read length is 41 or shorter. |
Whole Genome Sequencing v6.0 App Online Help
For Research Use Only. Not for use in diagnostic procedures.