DNA Sample Metrics Report
File name: DNA_SampleMetricsReport.txt
Review this report to assess DNA sample data quality using the following metrics.
|
Metric |
Description |
Passing Quality Threshold |
Indication When Failed |
|---|---|---|---|
|
Coverage_MAD |
Median of absolute deviations from the median of the normalized count of each CNV target region |
≤0.2 |
Less confidence in Copy Number Variant calling |
|
Median_BinCount_CNV_Targets |
Median of the raw read counts (normalized by region size) of each CNV target region |
≥1 |
Less confidence in Copy Number Variant calling |
|
Median_Insert_Size |
Median of fragment sizes calculated from read alignments |
≥79 |
Less confidence in Small Variant calling |
|
PCT_ExonBases_100X |
Percentage of exonic bases in gVCF that have greater than or equal to 100x coverage |
≥0.95 |
Less confidence in small variant calling |
|
PCT_Chimeric_Reads |
Percentage of unique reads that are part of a chimeric alignment (ie, one half of the read maps to one location and the other half maps to a nonsequential position) out of total mapped unique reads |
N/A |
Not used to evaluate sample data quality |
For Research Use Only. Not for use in diagnostic procedures.