Libraries prepared with custom sequencing primers require optimization to perform successfully on Illumina sequencing platforms. Combining (or “spiking”) custom primers into the Illumina primers is necessary when including PhiX or other standard Illumina libraries with the custom sample. Illumina cannot guarantee the performance or compatibility of custom primers, and it is up to the customer to validate primers for sequencing on a specific platform. When using custom sequencing primers, it is critical to know the recommended platform-dependent loading concentration, total volume, loading positions, and primer annealing temperatures. This bulletin provides information for spiking custom primers into the Illumina primers.
Spiking custom primers into Illumina sequencing primers
Important Note: The guidelines below are based on the current primer volumes. To ensure accuracy, measure total primer volumes in the cartridge with a pipette before proceeding with setup.
iSeq 100
With the release of iSeq Control Software version 3.0, we are now able to sequence with custom primers on the iSeq 100. Due to the construction of the iSeq 100 cartridge, it is not possible to spike custom primers directly into the pre-loaded Illumina primers.
In lieu of spiking custom primers directly into reagent cartridge wells, the NextSeq 1000/2000 custom primer kit configurations are compatible with the iSeq 100 and should be used. Custom primers should be added to the appropriate Illumina primers, then loaded into the iSeq 100 reagent cartridge using the custom primer workflow. For available kit configurations and their part numbers, please refer to the NextSeq 1000/2000 Custom Primers Guide.
Important Note: The NextSeq 1000/2000 Custom Primers Guide provides primer addition guidelines with volumes specific to the NextSeq 1000/2000. Please see below for iSeq 100-specific guidance for adding custom primers.
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (µL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| i1 v2 | Read 1 (HP21) | NA | 140 | 0.3 |
| Index 1 (i7) (BP14) | NA | 140 | 0.6 | |
| Index 2 (i5) (BP14) | NA | 140 | 0.6 | |
| Read 2 (HP21) | NA | 140 | 0.3 |
MiniSeq
Because of the unique dual-indexing workflow on the MiniSeq, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined here.
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (µL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| Rapid Run 100 Cycles |
Read 1 (BP10) | 24 | 590 | 0.3 |
| Read 2 (BP11) | 25 | 530 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 28 | 740 | 0.3 | |
| High Output 75 Cycles |
Read 1 (BP10) | 24 | 550 | 0.3 |
| Read 2 (BP11) | 25 | 610 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 28 | 820 | 0.3 | |
| High Output 150 Cycles |
Read 1 (BP10) | 24 | 550 | 0.3 |
| Read 2 (BP11) | 25 | 610 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 28 | 857 | 0.3 | |
| High Output 300 Cycles |
Read 1 (BP10) | 24 | 550 | 0.3 |
| Read 2 (BP11) | 25 | 610 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 28 | 820 | 0.3 | |
| Mid Output 300 Cycles |
Read 1 (BP10) | 24 | 550 | 0.3 |
| Read 2 (BP11) | 25 | 610 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 28 | 820 | 0.3 |
MiSeq
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (µL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| v2 and v3 | Read 1 (HP10) | 12 | 680 | 0.5 |
| Index 1 (i7) (HP12)* | 13 | 680 | 0.5 | |
| Read 2 (HP11) | 14 | 680 | 0.5 |
* There is no option for a custom Index 2 (i5) primer since the template uses the grafted P5 primer on the surface of the flow cell.
NextSeq 500/550
Because of the unique dual-indexing workflow on the NextSeq 500/550, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined here.
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (mL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| High Output 75 cycles |
Read 1 (BP10) | 20 | 1.72 | 0.3 |
| Read 2 (BP11) | 21 | 1.98 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 22 | 2.83 | 0.3 | |
| High Output 150 Cycles |
Read 1 (BP10) | 20 | 1.73 | 0.3 |
| Read 2 (BP11) | 21 | 1.98 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 22 | 2.83 | 0.3 | |
| High Output 300 Cycles |
Read 1 (BP10) | 20 | 1.73 | 0.3 |
| Read 2 (BP11) | 21 | 1.98 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 22 | 2.83 | 0.3 | |
| Mid Output 150 Cycles |
Read 1 (BP10) | 20 | 1.34 | 0.3 |
| Read 2 (BP11) | 21 | 1.51 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 22 | 2.09 | 0.3 | |
| Mid Output 300 Cycles |
Read 1 (BP10) | 20 | 1.33 | 0.3 |
| Read 2 (BP11) | 21 | 1.52 | 0.3 | |
| Index 1 (i7) and Index 2 (i5) (BP14) | 22 | 2.09 | 0.3 |
NextSeq 1000/2000
Due to the construction of the NextSeq 1000/2000 cartridge, it is not possible to load custom primers into the Illumina primers.
If Illumina primers are needed to sequence PhiX during a run, a NextSeq 1000/2000 Custom Primer kit is required. Refer to the NextSeq 1000/2000 Custom Primers Guide for instructions.
HiSeq 1000/1500/2000/2500
| Kit version | Illumina Primer (name) | Position | Total Volume (mL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| High Output v4 | Read 1 (HP10) | cBot plate - Row 2 | 0.3 per tube | 0.5 |
| Index 1 (i7) (HP12) | PE rack - 17 | 2.5 | 0.5 | |
| Index 2 (i5) (HP9) on a single read run* |
PE rack - 16 | 2.5 | 0.5 | |
| Read 2 (HP11) | PE rack - 16 | 2.5 | 0.5 | |
| High Output v3 | Read 1 (HP6) | cBot plate - Row 11 | 0.3 per tube | 0.5 |
| Index 1 (i7) (HP8 or HP12) | PE rack - 17 | 3.15 | 0.5 | |
| Index 2 (i5) (HP9) on a single read run* |
PE rack - 16 | 3.15 | 0.5 | |
| Read 2 (HP7 or HP11) | PE rack - 16 | 3.15 | 0.5 | |
| Rapid Run v2 | Read 1 (HP10) | PE rack - 18 | 1.93 | 0.5 |
| Index 1 (i7) (HP12) | PE rack - 17 | 1.93 | 0.5 | |
| Index 2 (i5) (HP9) on a single read run* |
PE rack - 16 | 1.93 | 0.5 | |
| Read 2 (HP11) | PE rack - 16 | 1.93 | 0.5 |
* There is no option for a custom Index 2 (i5) primer since the template uses the grafted P5 primer on the surface of the flow cell.
HiSeq 3000/4000
| Kit version | Illumina Primer (name) | Position | Total Volume (mL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| High Output | Read 1 (HP10) | cBot plate - Row 11 | 0.39 per tube | 0.5 |
| Index 1 (i7) and Index 2 (i5) (HP14) |
PE rack - 17 | 4.31 | 0.5 | |
| Read 2 (HP11) | PE rack - 16 | 2.64 | 0.5 |
NovaSeq
Because of the primer design and dual-indexing workflow on the NovaSeq, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined here.
NovaSeq v1.0 Consumables
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (mL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| SP 100/200/300/500 cycles |
Read 1 (BP10) | 24 | 4 | 0.3 |
| Index 1 (i7) (BP14)* | 23 | 5 | 0.3 | |
| Read 2 (BP11) | 13 | 2 | 0.3 | |
| S1 and S2 100/200/300 cycles |
Read 1 (BP10) | 24 | 4 | 0.3 |
| Index 1 (i7) (BP14)* | 23 | 5 | 0.3 | |
| Read 2 (BP11) | 13 | 2 | 0.3 | |
| S4 200/300 cycles |
Read 1 (BP10) | 24 | 7.3 | 0.3 |
| Index 1 (i7) (BP14)* | 23 | 5 | 0.3 | |
| Read 2 (BP11) | 13 | 3.5 | 0.3 |
* There is no option for a custom Index 2 (i5) primer since the template uses the grafted P5 primer on the surface of the flow cell.
NovaSeq v1.5 Consumables
| Kit version | Illumina Primer (name) | Cartridge Position |
Total Volume (mL) |
Custom primer final concentration (µM) |
|---|---|---|---|---|
| SP 100/200/300/500 cycles |
Read 1 (VP10) | 24 | 4 | 0.3 |
| Index 1 (i7) and Index 2 (i5) (VP14) | 23 | 5 | 0.3 | |
| Read 2 (VP11) | 13 | 2 | 0.3 | |
| S1 and S2 100/200/300 cycles |
Read 1 (VP10) | 24 | 4 | 0.3 |
| Index 1 (i7) and Index 2 (i5) (VP14) | 23 | 5 | 0.3 | |
| Read 2 (VP11) | 13 | 2 | 0.3 | |
| S4 200/300 cycles |
Read 1 (VP10) | 24 | 7.3 | 0.3 |
| Index 1 (i7) and Index 2 (i5) (VP14) | 23 | 5 | 0.3 | |
| Read 2 (VP11) | 13 | 3.5 | 0.3 |
Using Custom Primers in Place of Illumina Primers
If only the custom primer is needed, custom primers can be used in place of the Illumina primers. For instructions on how to set up a run using only custom primers, see the following guides:
For more information on the indexing strategies for each instrument, refer to the Indexed Sequencing Overview Guide.