Instrument maintenance and shutdown procedures for extended site closures

06/12/20


As we carefully monitor the emerging situation related to the outbreak of COVID-19 (Novel Coronavirus), this bulletin provides detailed instructions for shutting down and maintaining Illumina instruments in the event of lab closure. Read Illumina Statement on Novel Coronavirus and follow the Illumina Perspective on the Novel Coronavirus SARS-CoV-2 (2019-nCoV) Outbreak web page for further updates.

General Instructions | MiSeq | NextSeq 500/550 | MiniSeq | iSeq 100 | HiSeq Systems | NovaSeq | iScan | Tecan | Hamilton MICROLAB STAR | Shutting down the VeriSeq NIPT Server

General instructions for shutting instruments down for an extended period of time

General instructions for sequencing systems

  1. Perform a maintenance/manual/Standby wash (Standby wash for MiSeq only) using Tween 20 and NaOCl (if applicable), instructions are found in the instrument system guides
  2. Perform another maintenance/manual/Standby wash (Standby wash for MiSeq only) using laboratory-grade water to remove Tween from the system
  3. If possible, perform the above steps every 30 days

General instructions for all systems

  1. Shut down procedures:
    • If employees will be in the lab, leave instrument powered on while idle
    • If lab is completely shutting down (no employees on site or possibility of power being shut off), power off the instrument via normal processes in the instrument system guide
  2. Keep the instrument off until normal work schedules permit

Upon permanent return to work

  1. Power on the instrument
  2. Perform two maintenance/manual washes to rehydrate the fluidics system
    • Note: If additional maintenance/manual/Standby wash using laboratory-grade water was not performed prior to instrument shutdown, perform three maintenance/manual washes to rehydrate the fluidics system and run system checks to test the fluidics delivery

 


 

Instrument-specific instructions for Maintenance Best Practices during normal operation

MiSeq™

In the event of an extended laboratory shutdown, the MiSeq instrument can be placed into Standby mode by performing a Standby wash. When in Standby mode, perform additional Standby washes every 30 days that the instrument is idle. To switch the instrument out of Standby mode, perform a Maintenance wash. If the instrument has been idle for a significant period, additional Maintenance washes can be performed to ensure that the fluidics system is ready for a sequencing run.

Instructions for performing Standby and Maintenance washes are found in the MiSeq System Guide.

Shutdown the Instrument:

  1. Close MiSeq Control Software
  2. Make sure that no other programs are running, then shut down your PC from the Windows Start button
  3. After the PC has shut down, toggle the power switch at the back of the MiSeq to turn off the instrument

NextSeq™ 500/550

For the NextSeq 500/550, perform a Quick wash every 14 days that the instrument is idle, leaving the reagent wash cartridge and buffer cartridge in place.

Instructions for performing the Quick Wash are found on Page 36 of the NextSeq 500/550 System Guide.

Shutdown the Instrument:

  1. Select Manage Instrument
    NOTE: To shutdown the NextSeq 550Dx instrument in research mode, see NextSeq 550Dx Reboot and Shut Down Options on page 67 of the NextSeq 500/550 System Guide
  2. Select Shutdown Options
  3. Select Shutdown
  4. The Shutdown command safely shuts down the software and turns off instrument power
    CAUTION: Do not relocate the instrument. Moving the instrument improperly can affect the optical alignment and compromise data integrity. If instrument relocation is necessary, contact your Illumina representative.

MiniSeq™

For the MiniSeq instruments, perform a Quick wash every 7 days that the instrument is idle, leaving the reagent wash cartridge and buffer cartridge in place.

Instructions for performing the Quick Wash are found on Page 26 of the MiniSeq System Guide.

Shutdown the Instrument:

  1. Select Manage Instrument
  2. Select Shutdown Options
  3. Select Shutdown
  4. The Shutdown command safely shuts down the software and turns off instrument power
    CAUTION: Do not relocate the instrument. Moving the instrument improperly can affect the optical alignment and compromise data integrity. If instrument relocation is necessary, contact your Illumina representative.

iSeq™ 100

Because the fluidics system on the iSeq 100 is housed within the reagent cartridge, no additional steps are required for idling the iSeq 100. For extended closures, shut down the instrument by navigating to the hamburger menu in the top left corner of the Control Software and selecting Shutdown System. After the instrument has been shut down, toggle the power switch on the back of the instrument to fully power down the system.

More information on the iSeq 100 is found in the iSeq 100 System Guide.

HiSeq™ 2500/3000/4000/X

To idle the HiSeq instrument for 10 days or less, use the following instructions:

  1. Perform a maintenance wash to flush the system
  2. Leave the flow cell on the flow cell stage with the flow cell lever in position 2. Leave the manifolds in the raised position
  3. Load 10 ml laboratory-grade water in each position in the reagent racks, then lower the sippers
  4. Load 1 ml laboratory-grade water in the loading station position (HiSeq 2500 only)
  5. Before using the instrument, perform a water wash with laboratory-grade water

If the instrument will be idle for more than 10 days, shut down the instrument using the following instructions:

  1. Perform a maintenance wash to flush the system
  2. Remove the flow cell from the flow cell stage
  3. Wipe the surface of the flow cell holder using an alcohol wipe or a lint-free tissue moistened with ethanol or isopropanol
    CAUTION: Do not allow alcohol to drip into the vacuum holes or around the manifolds. Use a low-lint lab tissue to dry the stage, if necessary
  4. Load 10 ml laboratory-grade water in each position in the reagent racks, then lower the sippers
    • (HiSeq 2500 only) Load 1 ml laboratory-grade water in each loading station position
  5. Turn off the instrument

More information on HiSeq systems is found in the respective HiSeq System Guides found here:

NovaSeq™

The NovaSeq 6000 instrument requires a maintenance wash every 14 days that the instrument has been idle or if a four-lane run with an automatic post-run wash has not been completed within the past 14 days.

Instructions for performing the Maintenance wash on the NovaSeq 6000 instrument are found on Page 56 of the NovaSeq 6000 System Guide.

If employees will be in the lab, leave NovaSeq 6000 powered on while idle. If lab is completely shutting down, power off the NovaSeq 6000.

Shutdown the instrument:

  1. Choose Instrument Shutdown from NovaSeq Control Software hamburger menu
  2. When monitor shuts down, use the toggle switch at rear right to switch off the instrument

iScan™

Shutdown the iScan System:

  1. Remove BeadChips
    1. Open the iScan Reader tray
    2. To remove the carrier, lift it straight up and out of the tray
  2. Close the iScan Control Software
    1. Select the menu button in the upper left corner of the screen
    2. Select Exit
  3. From the Windows Start menu on the iScan System computer, select Shutdown
  4. Toggle the Power switch on the back panel of the iScan Reader

For more information on maintenance or shutting down the iScan instrument, refer to the iScan System Guide.

Tecan

Preparing the Tecan for extended idle period:

  1. Follow Cleaning the Tecan instructions from the Infinium Assay Lab Setup and Procedures Guide
  2. After washing with 10% bleach
    1. Fully remove the bleach solution from the carboy and refill with lab-grade water
    2. Perform two additional washes on the Tecan with lab-grade water
  3. Repeat above steps every 30 days if possible, otherwise leave the system powered off
  4. Powering off the Tecan
    • If lab personnel will be available, leave the Tecan powered on
    • If no lab/institution personnel will be available, power off the Tecan

Powering off the Tecan:

  1. Dry out the Tecan fluidics system
    1. Empty out the water tank
    2. Run two system washes to dry out the lines
  2. Power down the water circulator/Te-Flow
    1. Power down the water circulator
    2. Drain the water tubing to and from the Te-Flow
    3. Rinse the Te-Flow pan with water and let drain to waste
    4. Empty waste container
  3. Shut down the Illumina Automation Control (IAC) software
  4. Shut down the computer
  5. Power down the Tecan

Upon permanently returning to work:

  1. Power on the Tecan
  2. Refer to the following sections in the Infinium Assay Lab Setup and Procedures Guide under Robot Usage and Maintenance
    • Preparing the Robot for Use (perform at least two bleach washes)
    • Testing Volume Accuracy of the Tecan Tips

For more information on maintaining or shutting down the Tecan instrument, refer to the Infinium Assay Lab Setup and Procedures Guide.

Hamilton MICROLAB STAR

For the Hamilton MICROLAB STAR the weekly maintenance should be performed every 7 days while the instrument is idle.

  1. Launch the VeriSeq NIPT Maintenance and verification method from the Application Launcher.
  2. Select the desired maintenance routine by selecting the specific check box and by selecting the “Run Process” button. Instructions will be issued on-screen detailing all procedures required to perform the selected maintenance routine.

Shutting down the VeriSeq NIPT Server

Logging into the Web UI

  1. On a computer connected to the same network as the Onsite Server, open one of the following web browsers:
    • Chrome v69 or later
    • Firefox v62 or later
    • Internet Explorer v11 or later
  2. Enter the server IP address or the server name provided by Illumina at installation, equivalent to:
    \\[VeriSeq Onsite Sever v2 IP address]\login.
    For example, \\10.10.10.10\login.
  3. If a browser security warning appears, add a security exception to proceed to the login screen.
  4. In the login screen, enter the case-sensitive user name and password provided by Illumina, and select Log In.

Shutting down the server

Note: Only Service Engineers and Administrators have permission to shut down the server.

To shut down the Onsite Server:

  1. On the Settings drop-down list, select Shut Down Server.
  2. Select Shut Down to shut down the Onsite Server, or select Cancel to exit without shutting down.
  3. Enter a reason for shutting down the Onsite Server. The reason is logged for troubleshooting purposes.