TruSight Oncology 500 ctDNA Compatible Products

Compatible Sequencing Systems

NovaSeq 6000

Settings for Sequencing

Illumina recommends setting up a paired-end run with 151 cycles per read (2 × 151) and 8 cycles per index read.

Compatible Analysis Software

TSO 500 ctDNA Local App v1.0.1 (compatible with v1 sequencing consumables only)

TSO 500 ctDNA Local App v1.1 (compatible with both v1 and v1.5 sequencing consumables)

TSO 500 ctDNA Local App v1.2 (compatible with both v1 and v1.5 sequencing consumables)

Compatible Analysis Hardware

Illumina DRAGEN Bio-IT Platform v3.6 (DRAGEN On-Premise)

TSO 500 ctDNA/DRAGEN Compatibility Information

DRAGEN TruSight Oncology ctDNA Analysis Software License

DRAGEN TruSight Oncology ctDNA Analysis Software License. License quota is included in the library prep assay price.

Number of Libraries/Compatibility

For optimal coverage, use the following guidelines for the number of libraries per sequencing run.

NovaSeq 6000 Flow Cell Flow Cell Loading Minimum Number of Libraries Maximum Number of Libraries
S2 OBEM/Standard 6 8
S4 OBEM/Standard 16 16
S4 Xp 6 libraries per lane 6 libraries per lane