Unique Molecular Identifiers
DRAGEN can process data from whole genome and hybrid-capture assays with unique molecular identifiers (UMI). UMIs are molecular tags added to DNA fragments before amplification to determine the original input DNA molecule of the amplified fragments. UMIs help reduce errors and biases introduced by DNA damage such as deamination before library prep, PCR error, or sequencing errors.
To use the UMI Pipeline, the input reads files must be from a paired-end run. Input can be pairs of FASTQ files or aligned/unaligned BAM input.
DRAGEN supports the following UMI types:
• | Dual, nonrandom UMIs, such as TruSight Oncology (TSO) UMI Reagents or IDT xGen Prism. |
• | Dual, random UMIs, such as Agilent SureSelect XT HS2 molecular barcodes (MBC) or IDT xGen Duplex Seq Adapters. |
• | Single-ended, random UMIs, such as Agilent SureSelect XT HS molecular barcodes (MBC) or IDT xGen dual index UMI Adapters. |
DRAGEN uses the UMI sequence to group the read pairs by their original input fragment and generates a consensus read pair for each such group, or family. The consensus reduces error rates to detect rare and low frequency somatic variants in DNA samples with high accuracy. DRAGEN generates a consensus as follows.
1. | Aligns reads. |
2. | Groups reads into groups with matching UMI and pair alignments. These groups are referred to as families. |
3. | Generates a single consensus read pair for each read family. |
These generated reads have higher quality scores than the input reads and reflect the increased confidence gained by combining multiple observations into each base call.
UMI workflow is only compatible with small variant calling and SV in DRAGEN

Enter UMIs in one of the following formats:
• | Read name—The UMI sequence is located in the eighth colon-delimited field of the read name (QNAME). For example, NDX550136:7:H2MTNBDXX:1:13302:3141:10799:AAGGATG+TCGGAGA |
• | BAM tag—The UMI is present as an RX tag in prealigned or aligned BAM file (standard SAM format). |
• | FASTQ file—The UMI is located in a third FASTQ file using the same read order as the read pairs. |
To create FASTQ, append the UMI to the read name, and then specify the appropriate OverrideCycles setting in the DRAGEN BCL conversion tool (see DRAGEN BCL Data Conversion). DRAGEN supports UMIs with two parts each with a maximum of 8 bp and separated by +, or a single UMI with a maximum of 15 bp.

For dual, nonrandom UMIs, you can provide a predefined UMI correction table or a list of valid UMI sequences as input. To create the UMI correction table, use a tab-delimited file, include a header, and add the following fields.
Field |
Value |
---|---|
UMI |
The UMI sequence. For example, ACGTAC. |
IsValid |
Specify if the UMI sequence is valid. Enter either: TRUE or FALSE. |
NearestCodes |
Colon-separated list of nearest UMI sequences. For example, ACGTAA:ACGTAT. |
SecondNearestCodes |
Colon-separated list of second nearest sequences. For example, ACGGAA:ACGGAT. |
If customized correction table is not specified, DRAGEN uses the default table for TruSight Oncology (TSO) UMI Reagents. Alternatively, you can provide a file for whitelisted nonrandom UMI with valid UMI sequence one per line. DRAGEN then autogenerates a UMI correction table with hamming distance of one.

• | --umi-library-type—Set the batch option for different UMIs correction. Three batch modes are available that optimize collapsing configurations for different UMI types. Use one of the following modes: |
– | random-duplex—Dual, random UMIs. |
– | random-simplex—Single-ended, random UMIs. |
– | nonrandom-duplex—Dual, nonrandom UMIs. To use this option, provide a target manifest file using --umi-metrics-interval-file. |
• | --umi-min-supporting-reads—Specify the number of matching UMI inputs reads required to generate a consensus read. Any family with insufficient supporting reads is discarded. For example, the following are the recommended settings for FFPE and ctDNA. |
– | [FFPE] If the variant > 1%, use --umi-min-supporting-reads=1 with the --vc-enable-umi-solid variant caller parameter. For more information on variant caller options, see Variant Caller Options. |
– | ctDNA If the variant < 1%, use --umi-min-supporting-reads=2. with the --vc-enable-umi-liquid variant caller parameter. For more information on variant caller options, see Variant Caller Options |
• | --umi-enable—To enable read collapsing, set the --umi-enable option to true. This option is not compatible with –enable-duplicate-marking because the UMI pipeline generates a consensus read from a set of candidate input reads, rather than choosing the best nonduplicate read. If using the --umi-library-type option, --umi-enable is not required. |
• | --umi-emit-multiplicity—Set the consensus sequence type to output. DRAGEN UMI allows you to collapse duplex sequences from the two strands of the original molecules. Duplex sequence is typically ~20–60% of total library, depending on library kit, input material, and sequencing depth. Enter one of the following consensus sequence types: |
– | both—Output both simplex and duplex sequences. This option is the default. |
– | simplex—Output only simplex sequences. |
– | duplex—Output only duplex sequences. |
• | --umi-source—Specify the input type for the UMI sequence. The following are valid values: qname, bamtag, fastq. If using --umi-source=fastq, provide the UMI sequence from FASTQ file using --umi-fastq. |
• | --umi-correction-table—Enter the path to a customized correction table. By default, Local Run Manager uses lookup correction with a built-in table for the Illumina TruSight Oncology and Illumina for IDT UMI Index Anchor kits. |
• | --umi-nonrandom-whitelist—Enter the path for a customized, valid UMI sequence. |
• | --umi-metrics-interval-file—Enter the path for target region in BED format. |

DRAGEN processes UMIs by grouping reads by UMI and alignment position. If there are sequencing errors in the UMIs, DRAGEN can correct and detect small sequencing errors by using a lookup table or by using sequence similarity and read counts. You specify the type of correction with the --umi-correction-scheme option using the values “lookup,” “random,” or “none.”
For sparse sets of nonrandom UMIs, it is possible to create a lookup table that specifies which sequence can be corrected and how to correct it. This correct file scheme works best on UMI sets where sequences have a minimum hamming/edit distance between them. By default, DRAGEN uses lookup correction with a built-in correct table for the Illumina TruSight Oncology and Illumina for IDT UMI Index Anchor kits. Specify the path for your correction file using the --umi-library-type or --umi-correction-file option. If you are using a different set of nonrandom UMIs, contact Illumina Technical Support for information on generating the corresponding correction file.
In the random UMI correction scheme, DRAGEN must infer which UMIs at a given position are likely to be errors relative to other UMIs observed at the same position. The error modes include small UMI errors, such as one mismatch or UMI jumping or hopping artifact from library prep. DRAGEN accomplishes this as follows.
• | Groups reads by fragment alignment position. |
• | Within a small fuzzy window at each position, groups the reads first by the exact UMI sequence, which forms a family. |
• | Estimate UMI jumping or hopping probability through insert size distribution and number of distinct UMI at certain positions. |
• | Within a fuzzy window, calculates pair-wise likelihood ratio to assess if two families with different UMI sequences and genomic positions are derived from same original molecule. |
• | Merges families with likelihood lower than threshold. The default threshold is 1. |

Duplex UMI adapters simultaneously tag both strands of double-stranded DNA fragments. It is then possible to identify reads resulting from amplification of each strand of the original fragment.
Consumable Prefix considers two collapsed read pairs to be the sequence of two strands of the same original fragment of DNA if they have the same alignment position (within a fuzzy window), complementary orientations, and their UMIs are swapped from Read 1 and Read 2. If there is only single-ended UMI, DRAGEN compares the start-end position of families from two strands and computes pair-wise likelihood to determine if they are likely originated from two distinct families or should be merged as a duplex sequence. By default, DRAGEN outputs both simplex and duplex consensus sequences. To change the consensus sequence output type, use --umi-emit-multiplicity.

The following are example UMI commands:

The following is an example DRAGEN command for generating a consensus BAM file from input reads with Illumina UMIs:
dragen \
-r <REF> \
-1 <FQ1> \
-2 <FQ2> \
--output-dir <OUTPUT> \
--output-file-prefix <PREFIX> \
--enable-map-align true \
--enable-sort true \
--umi-enable true \
--umi-correction-scheme=lookup \
--umi-min-supporting-reads 2

To run with other random UMI library types, change --umi-library-type to random-simplex or random-duplex.
dragen \
-r <REF> \
-1 <FQ1> \
-2 <FQ3> \
--umi-source=fastq \
--umi-fastq <FQ2> \
--output-dir <OUTPUT> \
--output-file-prefix <
PREFIX> \
--enable-map-align true \
--enable-sort true \
--umi-library-type nonrandom-duplex \
--umi-metrics-interval-file [valid target BED file]

dragen \
-r <REF> \
-1 <FQ1> \
-2 <FQ2> \
--umi-correction-table <valid umi correction table> \
--output-dir <OUTPUT> \
--output-file-prefix <PREFIX> \
--enable-map-align true \
--enable-sort true \
--umi-library-type nonrandom-duplex \
--umi-metrics-interval-file <valid target BED file>

Consumable Prefix outputs an <output_prefix>.umi_metrics.csv file that describes the statistics for UMI collapsing. This file summarizes statistics on input reads, how they were grouped into families, how UMIs were corrected, and how families-generated consensus reads. The following metrics can be useful when tuning the pipeline for your application:
• | Discarded families—Any families having fewer than --umi-min-supporting-reads input or having a different duplex/simplex status than specified by --umi-emit-multiplicity are discarded. These reads are logged as Reads filtered out. The families are logged as Families discarded. |
• | UMI correction—Families may be combined in various ways. The number of such corrections are reported as follows. |
– | Families shifted—Families with a very closefragment alignment coordinates (up to the distance specified in the umi-fuzzy-window-size parameter). The default umi-fuzzy-window-size parameter is 3. |
– | Families contextually corrected—Families with the same fragment alignment coordinates and compatible UMIs are merged. |
– | Duplex families—Families with close alignment coordinates and complementary UMIs are merged. |
When you specify a valid path for --umi-metrics-interval-file, Consumable Prefix outputs a separate set of on target UMI statistics that contains only families within the specified BED file.
If you need to analyze the extent to which the observed UMIs cover the full space of possible UMI sequences, the histogram of unique UMIs per fragment position metric may be helpful. It is a zero-based histogram, where the index indicates a count of unique UMIs at a particular fragment position, and the value represents the number of positions with that count.
The following table describes available UMI metrics.
Metric |
Description |
---|---|
Number of reads |
Total number of reads. |
Number of reads with valid or correctable UMIs |
Number of reads for which the UMIs could be corrected based on the lookup table. |
Number of reads in discarded families |
Number of reads in discarded families. Families are discarded when there are not enough raw reads to support the family. |
Reads with all-G UMIs filtered out |
Number of reads filtered out due to all-G in UMI sequence. |
Reads filtered out |
Number of reads filtered out in total either for properties or in a discarded family. |
Reads with uncorrectable UMIs |
Number of reads where the UMI could not be corrected. |
Total number of families |
Number of simplex collapsed reads, |
Families contextually corrected |
Number of families that have some contextual correction. UMI correction is based on other families at the same mapping location. |
Families shifted |
Number of families that have some shift correction (merging of UMI families with very similar UMI and very close mapping locations. |
Families discarded |
Number of families filtered out by failing min supporting reads criteria or umi-emit type of simplex/duplex. |
Families discarded by min-support-reads |
Number of families filtered out by failing min supporting reads criteria. |
Families discarded by duplex/simplex |
Number of families filtered out by failing umi-emit type of simplex/duplex. |
Duplex families |
Number of families that are merged as duplex (both strands). For percentage, the denominator is the number of consensus pairs. |
Consensus pairs emitted |
Number of collapsed reads in output BAM. |
Mean family depth |
Average number of reads per family |
Histogram of num supporting fragments |
Number of families with zero raw reads, one raw read, two raw reads, three raw reads, etc |
Number of collapsible regions |
Number of regions. |
Min collapsible region size (num reads) |
Number of reads in the most populated region. |
Max collapsible region size (num reads) |
Number of reads in the least populated region |
Mean collapsible region size (num reads) |
Average number of reads per region. |
Collapsible region size standard deviation |
Standard deviation of the number of reads per region. |
On target number of reads |
Number of reads that overlapped with UMI target interval --umi-metrics-interval-file. |
On target number of reads with valid or correctable UMIs |
Number of reads with a UMI that matched a UMI in the lookup table, including error allowance, and overlapped with UMI target interval. |
On target duplex families |
Number of families that are merged as duplex among all the families that are overlapped with UMI target interval. For percentage, denominator is the number of on-target consensus pairs. |
On target mean family depth |
Average number of reads per family that overlapped with UMI target interval. |
Histogram of unique UMIs per fragment position |
Number of positions with zero UMI sequences, one UMI sequence, two UMI sequences, etc. |
Total families in probability model estimation |
Total number of families used in estimation of UMI jumping rate and fragment size distribution used for probabilistic family merging. |
Number of potential Jumping Families |
Total number of families that are potential UMI jumping candidates and the corresponding ratio. |