06/16/2022
A ‘Bit error check failed’ error can indicate a transient communication issue with the instrument control software (Figure 1). The error can occur when launching the instrument control software on the NextSeq 500/550 or MiniSeq or when performing the pre-run check. Usually the error can be resolved by power cycling the instrument.

Figure 1. "Bit error check failed Originally caused by: DMA Read failed with return code 32"
To Power Cycle the Instrument:
After the power cycle is complete and the control software is initialized, repeat run setup and perform the pre-run check. If the pre-run check passes without error proceed with the sequencing run. No further action is required. If the issue persists, contact Illumina Technical Support for additional assistance.
Additional Resources: