Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
| Instrument | Impacted Software Versions | Actions Required |
|---|---|---|
| iScan | iScan Control Software 4.0.0 | Configure UCS account credentials. NOTE: The instructions indicate that iScan Control Software v4.0.0 and above require a software update. Version 4.0.0 does not need a software update and only requires configuration of UCS account credentials. |
| iScan Control Software 4.0.5 | Contact Illumina Technical Support to upgrade to v4.2.1. | |
| iSeq 100 | All | Configure UCS account credentials. |
| MiniSeq | MiniSeq Control Software 2.0 and newer | Configure UCS account credentials. |
| MiSeq | MiSeq Control Software 4.0 and newer | Configure UCS account credentials. |
| MiSeqDx | MiSeq Control Software 4.0 (RUO Mode) | Configure UCS account credentials. |
| MiSeqDx Operating Software 4.0.1 and newer |
|
|
| NextSeq 500/550 | NextSeq Control Software 4.0 | Configure UCS account credentials. |
| NextSeq 550Dx | NextSeq Control Software 4.0 (RUO Mode) | Configure UCS account credentials. |
| NextSeq Operating Software 1.0.0-1.3.1 | Configure UCS account credentials. | |
| NextSeq Operating Software 1.3.3 and newer |
|
|
| NextSeq 1000/2000 | NextSeq Control Software 1.4.1 | Update the system software with NextSeq 1000/2000 Control Software Suite v1.5.0 Installer. |
| NovaSeq 6000 | NovaSeq Control Software 1.7 and older | Configure UCS account credentials. NOTE: The instructions state that the firewall port must be closed for these software versions. Closing the firewall port is not required. |
| NovaSeq Control Software 1.8 |
|
The following video is meant to provide additional context to the UCS Instruction Guide. Please read the instructions first.
This video:
NOTE: This video is applicable only to instruments where UCS is currently configured to use "Local System". If UCS is not configured to use "Local System" on your instrument, refer to the Instruction Guide for additional details.