View Run & Lane Metrics
The Run & Lane Metrics page has the overall statistics about the run. Use this option when you want to view information about the run, such as percent alignment, cycles, and densities.
Select the Metrics tab from a run page to view metrics.
The following metrics appear in the top table, split out by read and total.
|
Metrics |
Description |
|---|---|
|
Cycles |
The number of cycles in the read. |
|
Yield |
The number of bases sequenced, which is updated as the run progresses. |
|
Projected Yield |
The projected number of bases expected to be sequenced at the end of the run. |
|
Aligned |
The percentage of the sample that aligned to the PhiX genome, which is determined for each level or read independently. |
|
Error Rate |
The calculated error rate of the reads that aligned to PhiX. |
|
Intensity Cycle 1 |
The average of the A channel intensity measured at the first cycle averaged over filtered clusters. For the NextSeq 500 System, the red channel is used. |
|
%Q ≥ 30 |
The percentage of bases with a quality score of 30 or higher, respectively. This chart is generated after the 25th cycle, and the values represent the current cycle. |
The following metrics are available in the Read tables, split out by lane.
|
Metrics |
Description |
|---|---|
|
Clusters PF |
The percentage of clusters passing filtering, +/- 1 standard deviation. |
|
%Q ≥ 30 |
The percentage of bases with a quality score of 30 or higher, respectively. This chart is generated after the 25th cycle, and the values represent the current cycle. |
|
Yield |
The number of bases sequenced which passed filter. |
|
Tiles |
The number of tiles per lane. |
|
Error Rate |
The calculated error rate, as determined by the PhiX alignment. Select Subsequent columns display the error rate for cycles 1–35, 1–75, and 1–100. |
|
Error Rate @cycle |
The error rate for cycles 1–35, 1–75, and 1–100. Select the Error Rate value to view this metric. |
|
Cycles Err Rated |
The number of cycles that have been error-rated using PhiX, starting at cycle 1. Select the Error Rate value to view this metric. |
|
Aligned |
The percentage that aligned to the PhiX genome. Select the Error Rate value to view this metric. |
|
Reads PF |
The number of clusters (in millions) passing filtering. |
|
Total Reads |
The number of clusters (in millions). Select the Reads PF value to view this metric. |
|
Density |
The density of clusters (in thousands per mm2) detected by image analysis, +/- 1 standard deviation. |
|
Phas./Prephas. |
The value used by RTA for the percentage of molecules in a cluster for which sequencing falls behind (phasing) or jumps ahead (prephasing) the current cycle within a read. For MiSeq |
|
Phasing Slope/Offset, Prephasing Slope/Offset |
The best-fit slope and offset of the phasing/prephasing corrections, calculated from the entire read. |
|
Intensity |
The average of the A channel intensity measured at the first cycle averaged over filtered clusters. |
|
Status |
The QC status for the run. Select the QC status to view detailed QC results. |