DNA Library QC Metrics for Small Variant Calling
Metric |
Description |
Recommended Threshold |
---|---|---|
PCT_CHIMERIC_READS (%) |
Percentage of unique reads that are part of a chimeric alignment (for example, half of the read maps to one location and the other half maps to a nonsequential location) out of the total number of mapped unique reads. |
≤ 3.5 |
MEDIAN_INSERT_SIZE (bp) |
Median of fragment sizes calculated from read alignments. |
≥ 70 |
PCT_EXON_250X (%) |
Percentage of exonic bases in gVCF that have greater than or equal to 250X coverage. |
≥ 81.3 |