Data By Cycle

Data by Cycle graphs the progression by cycle of quality metrics during a run.

Data By Cycle Showing Intensity

The Data by Cycle chart has the following features:

• You can select the displayed metric, surface (if your system scans multiple surfaces), cycle, and channel (base) using the drop-down lists.
• The chart is displayed with tailored scaling by default, or you can fix the Y-axis scale by selecting the Fix Scale checkbox.
• Some metrics (% ≥Q20 and % ≥Q30) are monitored for a single cycle by default, or you can monitor the cumulative metrics (up to that cycle) by selecting the Accum checkbox.
• You can focus on an area of interest by dragging to pan the view or using the mouse wheel to zoom in.

The following table details the possible metrics displayed in this chart. The available options vary by run.

Data By Cycle Chart Options

Setting

Description

Intensity

This chart shows the intensity by color and cycle of the 90% percentile of the data for each tile.

FWHM

The percentage of clusters for which the selected base (A, C, T, or G) has been called.

% Base

The percentage of clusters for which the selected base (A, C, T, or G) has been called.

% NoCall

The percentage of clusters that have no call.

% ≥Q20 and % ≥Q30

The percentage of bases with a quality score of > 20 or > 30, respectively. These charts are generated after the 25th cycle, and the values represent the current scored cycle.

Median Q-Score

The median Q-score for each tile over all bases for the current cycle. These charts are generated after cycle 25. Use this setting to examine the Q-scores of your run as it progresses. Because it relies on a single threshold, the %≥Q30 plot can be oversimplified.

Error Rate

The calculated error rate, as determined by a spiked in PhiX control sample. If no PhiX control sample is run in the lane, this chart is not available.

Phasing, Prephasing

The estimated percentage of molecules in a cluster for which sequencing falls behind (phasing) or jumps ahead (prephasing) the current cycle within a read.

Legacy Phasing Rate, Legacy Prephasing Rate

The estimated percentage of molecules in a cluster for which sequencing falls behind (phasing) or jumps ahead (prephasing) the current cycle within a read.

Legacy rates consider only the first 25 cycles.

Corrected Intensity

The intensity corrected for cross talk between the color channels by the matrix estimation and phasing and prephasing.

Called Intensity

The intensity for the called base.

Signal to Noise

The signal to noise ratio is calculated as mean called intensity divided by standard deviation of noncalled intensities.